| Integral to the Flow Transducer |
Accuracy: +/- .50% of Full Scale, [or 2.0% of reading whichever is better. Full Scale over the calibrated range. [Process fluid temperature span +/- 50°F] | Remote to 4000 ft., Wall or Frame mount |
| |
Repeatability: +/- 0.2 % of Rate | Signal Cable for Remote Design: 18 gauge, 2 pair with cable shield. Maximum distance, 4000 ft. |
| |
Response Time: | MODEL NUMBERING SYSTEM: |
Gases: 1 to 2 sec. Typical | Base Model: = 9500 |
Liquids: Less than 500ms | |
| |
Pressure Rating: | Electronics Mounting: |
0 – 1200 psig (Std) | I = Integral Mount |
0 – 10000 psig Dependent on meter size | R = Remote Mount |
| F = Fluid |
Flow Rates: | G = Gas |
Gas: 40 to 86,000 SFPM | L = Liquid |
Liquid: .02 to 20 FPS | |
Note: maximum determined by pipe capacity | |
| |
Flow Range: 100:1 Std. Multiple ranges available | Material: |
| 316SS = STD |
Calibrated Temperature Range: | Corrosion Resistant Materials available by request: Hastelloy C, Monel, Inconel, Tantalum, and Carpenter 20 |
Gas: -20 F to 350 F (-28°C to 162°C )STD. | Available Coatings: Carbide, Fluorocarbon , Teflon, and Sulfinert |
[-20°F to 600°F] (-28°C to 315°C) Optional) | Process Connection: |
Liquid: -20F to 350F (-28°C to 162°C ) STD. | PG = Packing Gland (Std.) |
[ -20°F to 500°F](-28°C to 260°C) Optional | FLG-150 =150# RF Flange (Std.Van stone) |
| TC = Triclover |
Wetted Materials: 316SS (STD). Consult the factory for other material requirements | FLG-Other = Consult Factory |
| Other = Consult Factory |
Connections: Flange, Packing Gland, Triclover, Hot Tap | |
| Power: |
Calibration Temperature Capability: -20°F to 600°F(-29°C to 315°C) | 24 = 22-26 VDC, 1 AMP. 4 WIRE |
| 120 = 120VAC, 1/2 AMP, 50/60 HZ (STD.) |
MICRO-CONTROLLER SPECIFICATIONS: | 240 = 240VAC 50/60 HZ |
16 BIT A/D | |
| Output Signals: |
Displays: Bright LED 2 Line 8 digit | (Select One**) |
Capable of displaying Current Flowrate and Total Flow (Optional) | 0/5 = 0-5 VDC** |
| 0/10 = 0-10 VDC** |
Communicator: | 4/20 = 4-20 mADC** into 750 ohms, Isolated std, Self Powered |
Interface: RS-232 via Ribbon Connector | 4/20T = Opt Process Fluid Temp. 4-20mA Self Powered |
Baud Rate: 9600 | Other = Consult Factory |
Filter Factor | |
Flow Full Scale | Display: (Optional) |
Flow Rate Decimal Point | ND = No Display |
Flow Totalizer Decimal Point | D = Flowrate Display, 8 Digit |
Flow Totalizer Flowrate Ratio | 8T = Totalizer |
| D/8T = Flowrate and Total Display, 8 digit |
Power: | |
120/240 vat, 50/60 Hz. | Signal Cable for Remote Design: SC25 = 25 FT Supplied with remote configuration |
24VDC, 1 Amp (Factory Set) | |
| |
Electrical Connections:3/4” FNPT Power & Signal | |
| |
Enclosure: Cast Aluminum, Epoxy Painted | |
| |
Environmental Ratings: | |
Housing: Explosion Proof, Class 1, Div. 1, Groups, B, C D | |
Weatherproof: Nema 4 | |
Ambient Temperature: -25 to 60 C (-13 to 140 F) | |
Storage Temperature: -40 to 70 C ( -40 to 158F ) |